• The cross cylinder setting cannot be selected between Flipped and Auto. Set the lens type
with the “Cross cylinder (XC) test” parameter (see page 134).
• The fog function can be added with both eyes open. Set the fog according to "3.8 Fog
Function with Both Eyes Open" (page 115). The setting is programmed the same as other
auxiliary lenses.
• When the “Intelligent XC” parameter is set to “Yes” (default), the cross cylinder test may not
be performed depending on the cylinder values (cylindrical power).
125
ADVANCED FEATURES
Programming
3
For Plus Package, exam can be specified with the 21 Point Exams.
1) Press [Chart] to display the 21 Point
Exam button.
2) Select the exam to be stored with the
21 Point Exam button.
21 Point Exams button
[#7 Subj (F)] cannot be selected.
The contents of exam set (chart, auxiliary
lens etc.) are written.
6
When the steps on the screen are pro-
grammed, go to the next page.
Press [
Next].
When all steps are programmed, press
[
].
To clear the steps from the next to last for
overwriting, press [LastTest].
7
Repeat Steps 4 to 6 until the desired charts are programmed in the order desired to be
measured.
It is possible to store up to 32 steps.
8
Exit from programming mode.
Press [
] until the screen returns to the normal measurement screen.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...