141
ADVANCED FEATURES
Parameter Settings
3
Preset power of Subj 2: All data, CYL=0, Clear
Factory setting: All data
Selection of whether or not to use the data entered in the Preset power of Subj 1 parameter without
being corrected.
All data
Uses all data without being corrected.
CYL=0
Uses only sphere data. (The cylindrical and axis values become 0.)
Clear
Enters 0 in all data fields.
This function is not available for the data entered with the dial.
Copy LM/AR data when switching from Day to Night: Yes, No
Factory setting: No
Selection of whether or not to copy the day data to the night LM field or AR field that is not entered
when Day mode is switched to Night mode. This parameter can be selected only when the RT is con-
nected with a chart presenting device equipped with the night vision test function.
Intelligent XC: Yes, No
Factory setting: Yes
When the setting is Yes in the cross cylinder test, the test contents change depending on the cylinder
values. It allows the test to be performed easily.
CYL=0: The message whether the cross cylinder test and previous red-green test are skipped or
not appears.
CYL=0.25: The cross cylinder test is performed in the order of cylindrical power
cylindrical axis
cylindrical power.
When the cylinder value is 0 in the first cylindrical power measurement, the next cross cylinder
test is not performed.
CYL
0.5: The cross cylinder test is performed in the order of cylindrical axis
cylindrical power.
When the setting is No, the cross cylinder test is performed in the programmed order regardless of the
cylinder value.
Shift + Final: Final Fit, copy the data
Factory setting: Final Fit
Setting for the function when Final is pressed together with SHIFT.
Final Fit: Enters the far point auto adjustment mode. See "2.7.2 Auto adjustment of far powers (with
Program B)" (page 84) and "2.7.3 Power adjustment without standard programs" (page 85).
copy the data: Copies the displayed numeric value to the prescription field regardless of the presence
or absence of prescription mode data.
Dial switch: S
C
A, S
C
A
VA, S
A
C, S
A
C
VA
Factory setting: S
C
A
Setting for the function with the dial switch.
Each time it is pressed, the modes change in the selected order.
Lens for Retino: +1.50 D, +2.00 D
Factory setting: +1.50 D
Setting for the spherical lens for retinoscope.
Distance between the refractor head and retinoscope
+1.50 D: 67 cm
+2.00 D: 50 cm
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...