84
OPERATING PROCEDURE
Power Adjustment (Final Fit)
2
Select [High] or [Low].
The auto adjustment function calculates
the Final Fit value.
The bar graph appears during calculation.
When calculation is completed, the Final
Fit value is displayed as Final 1 and cor-
rective lenses are placed in the measuring
windows. The chart of 20/20 to 20/10 is
presented automatically.
The LM and subjective (monocular full cor-
rection) data are displayed in the left and
right windows. It makes comparison easier.
2.7.2
Auto adjustment of far powers (with Program B)
Program B does not contain the auto adjustment function. Perform the following procedures to use the
function.
* The “SHIFT+Final” parameter must be set to “Final Fit”.
1
Measure the binocular full correction, press
after the stereo test.
The message whether to use the auto
adjustment function appears.
2
Press
while pressing
.
*18
The message whether to enter the LM
data (glasses data) appears.
When the data has already been entered,
go to Step 4.
*18.
When the auto adjustment function is not used, adjust the powers manually and press
.
161221687130 承認済
Uncontrolled copy when printed.
Содержание RT-5100
Страница 12: ...X 161221687130 承認済 Uncontrolled copy when printed ...
Страница 18: ... XVI 161221687130 承認済 Uncontrolled copy when printed ...
Страница 46: ...28 BEFORE USE Getting Started and Exiting 161221687130 承認済 Uncontrolled copy when printed ...
Страница 226: ...208 TEST METHOD Visual Acuity Test by ETDRS chart 161221687130 承認済 Uncontrolled copy when printed ...
Страница 274: ...256 MAINTENANCE List of Replacement Parts 161221687130 承認済 Uncontrolled copy when printed ...
Страница 302: ...284 EMC ELECTROMAGNETIC COMPATIBILITY 161221687130 承認済 Uncontrolled copy when printed ...