1
Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
Thermo Scientific
TSQ 9000 Hardware Manual
55
23. Let the ion source cartridge cool down before removing it from the source holder.
Disassembling the Ion Source Cartridge
1. Disassemble the ion source cartridge by removing the locking ring first, then the repeller
spring, then the nut, insulator, and repeller (which comes out in one piece), ion volume,
lens 1, lens 2, and lens 3/RF lens.
2. Set the components on a clean work surface.
WARNING BURN HAZARD:
The ion source cartridge may be hot, so use caution.
ATTENTION RISQUE DE BRÛLURE
: soyez prudent, car la cartouche de la source d’ions
peut être chaude.
IMPORTANT
Many nitrile and latex gloves not certified for clean room use contain
silicone mold releasing agents that will contaminate the instrument. For this reason,
clean room gloves are strongly recommended when handling the ion source cartridge.
We recommend Cardinal Health CP100 Nitrile Cleanroom Gloves. See the
ISQ and
TSQ GC-MS Spare Parts Guide
for ordering information.
Figure 46.
Ion Source Cartridge Components
Ion Volume
Locking Ring
Repeller
Lens 1
Lens 2
Lens 3/RF Lens
Ion Cartridge Sleeve
Ion Volume
Repeller
Spring
Repeller
Nut
Ion
Volume-Repeller
Insulator