![Thermo Scientific TSQ 9000 Hardware Manual Download Page 361](http://html.mh-extra.com/html/thermo-scientific/tsq-9000/tsq-9000_hardware-manual_1104856361.webp)
3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
Thermo Scientific
TSQ 9000 Hardware Manual
333
8. Remove the controller interface board and PC communication board from the
instrument.
9. If you need to replace the support bracket, use a T10 Torxhead screwdriver to remove the
seven M3 screws securing the controller interface board to the support bracket.
10. Replace the bracket and reattach it to the controller interface board.
Figure 357.
Removing the Controller Interface Board Support Bracket
Support Bracket
Controller
Interface Board
PC Communication Board
Summary of Contents for TSQ 9000
Page 1: ...TSQ 9000 Mass Spectrometer Hardware Manual 1R120618 0003 Revision C December 2018 ...