3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
348
TSQ 9000 Hardware Manual
Thermo Scientific
Figure 373.
Disconnecting the Distribution Board
31. Disconnect all the cables from the distribution board. Also remove the cable that enters
the chassis from the back panel.
32. Remove the two flex cables that attach to the inlays on the front door.
33. Remove all 19 T10 Torxhead screws that connect the distribution board to the bottom of
the chassis.
34. Tilt the distribution board to one side and lift it out of the instrument.
35. Insert a new distribution board into the instrument and reattach the screws.
36. Reconnect the 1/16 in. collision gas tubing from the bulkhead compression fitting
located on the left hand sub panel.
37. Follow the steps in the
section to reattach the CI reagent
gas module and reconnect the inputs.
38. Reconnect all the cables.
39. Reattach the dynode/multiplier power supply to the back the chassis by reattaching the
screw.
40. Slide the main power supply back down into its slot.
41. Reattach the two screws holding the power supply to the instrument.
42. Reattach the power connectors to the inputs on the left hand sub panel.
Cables
Distribution Board