Advanced Troubleshooting
Replacing Fans and Filters
392
TSQ 9000 Hardware Manual
Thermo Scientific
Front Wire Mesh Filter
To replace the front wire mesh filter
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Use a T10 Torxhead screwdriver to remove the six screws holding the front wire mesh
filter to the interior front panel.
3. Remove and replace the front wire mesh filter.
4. Reattach the six screws holding the front wire mesh filter to the interior front panel.
Figure 417.
Replacing the Front Wire Mesh Filter
5. Restart your system by following the instructions in
Note
If needed, you can order an additional dust filter. See the
ISQ and TSQ GC-MS
Spare Parts Guide
for ordering information.
Note
You do not have to remove the ion source cartridge or column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
Screws