Advanced Troubleshooting
Replacing Components of the Analyzer
148
TSQ 9000 Hardware Manual
Thermo Scientific
Figure 113.
Removing the Analyzer Tray Screws
Figure 114.
Removing the Collision Cell and Quadrupole 3 from the Analyzer Tray
8. Disconnect the Q1 entrance lens wire from the source interface board.
9. Remove the two screws on each side of the Q1 entrance endcap.
Figure 115.
Removing Screws from the Quadrupole Rod Endcaps
Q1 Exit Endcap
Screw
Q3 Exit Endcap
Screw
Ion Guide Screws
Q3 Entrance Endcap Screw
Collision Cell
Assembly
Q3 Exit Lens
Assembly
Q3 Quad
Assembly
Screws
Collision Cell
Q1 Entrance
Lens and Endcap
Screws