3
Advanced Troubleshooting
Replacing Components of the Analyzer
164
TSQ 9000 Hardware Manual
Thermo Scientific
Quad 1 and Quad 3 Wire Sets
To replace the quad 1 or quad 3 wire sets
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the TSQ 9000 Instrument”
3. For both quad assemblies, use a T10 Torxhead screwdriver to remove the four screws
around the rods. These screws hold the wires in place. If the screws won’t come off the
rods, you must replace the whole quad assembly.
4. Use on the screws provided. These screws are made from a special material designed not
fuse with the quadrupole rods.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Figure 137.
Replacing the Q1 Quad Assembly Wires
Screws
Screws