25. Attach the source-repeller thumbscrews to the endcaps and twist them back on. When
you are tightening the thumbscrews, make sure the spacers stay in the groove or you could
crush them.
26. Reconnect the lens heater block grounding strap to the lens heater block
27. Reconnect the Q1 entrance lens, Q2 entrance lens, and Q3 exit lens wires to the source
interface board using a flat-tipped screwdriver.
Figure 156.
Reattaching the Source-Repeller Thumbscrews
Figure 157.
Reconnecting the Lens Heater Block Grounding Strap
Source-Repeller Thumbscrews
Lens
Heater
Block
Lens Heater
Grounding Strap
Screw
Summary of Contents for TSQ 9000
Page 1: ...TSQ 9000 Mass Spectrometer Hardware Manual 1R120618 0003 Revision C December 2018 ...