3
Advanced Troubleshooting
Replacing Components of the Ion Source
206
TSQ 9000 Hardware Manual
Thermo Scientific
13. Pull the vacuum interlock handle down to close the interior ball valve.
14. Twist the end of the vacuum interlock handle and flip the swing handle up so that it rests
next to the vacuum interlock knob location. Then twist the end again to hold it in place.
15. Replace the plug in the VPI to prevent accidental venting and excessive leaks.
16. Remove the tool from the instrument.
Figure 194.
Pulling Down the Vacuum Interlock Handle
CAUTION INSTRUMENT DAMAGE:
Do not move the main interlock handle when you
move the swing handle to the up position or you will open the interior ball valve, which
might cause you to lose vacuum and possibly damage the instrument.