Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
Thermo Scientific
TSQ 9000 Hardware Manual
305
Front and Rear RF Boards and Front and Rear Rod Driver Boards
To replace the front and rear radio frequency (RF) boards and front and rear rod drive
boards.
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. The front RF board is the second board standing on end toward the center of the
instrument. The rear RF board is the third board standing on end toward the center of
the TSQ 9000 instrument. Use a T20 Torxhead screwdriver to remove the screw holding
the board and bracket to the chassis.
4. Locating the RF Boards and Brackets
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering an
RF board. This board must be replaced by a Field Service Engineer, so when you place an
order for a new board, schedule the Field Service Engineer as well.
Note
You do not have to remove the ion source cartridge or column or vent the system. If
you are using hydrogen as a carrier gas, power off the GC.
Note
The steps to remove and replace the front and rear RF boards are the same. Refer to
the images for the proper location of each board.
CAUTION - INSTRUMENT DAMAGE
Limit handling of the boards as much as possible.
Do not touch any board components. Hold the board by the sheet metal bracket only.
ATTENTION DOMMAGES À L’INSTRUMENT
Limitez au maximum la manipulation des
cartes. Ne touchez aucun composant de carte. Gardez la carte uniquement près du support
en tôle.