7. Insert a new RF board/rod driver board kit by aligning the connectors on the bottom of
the board with the connectors on the distribution board beneath it. It will drop into the
slot when it is in the right position.
8. Push the board down into the connectors.
9. Gently push the board toward the instrument, carefully aligning the board over the
feedthrough pins, until it stops.
10. Reattach the screw holding the RF board to the bracket.
11. Reattach the top cover panel and tighten the screw holding it in place.
12. Restart your system by following the instructions in
Figure 334.
Reinstalling the RF Board/Rod Driver Board Kits
Note
This screws is required for good ground contact.
Note
The new RF board/rod driver board kit will have a slightly different performance
than your current board. It will need to be adjusted by a Field Service Engineer because
the adjustments must be made while the system is running.
Front RF
Board
Connectors
Rear RF
Board
Connectors