1
Performing Routine Maintenance
Replacing a Dual Filament
Thermo Scientific
TSQ 9000 Hardware Manual
77
14. Remove the filament.
15. Insert the replacement filament into its slot, rotate the spring into position, and slightly
tighten the screw holding the filament retaining clip in place.
16. Bend the filament wires so they do not touch any metal component. You may want the
old filament as a model.
17. Attach the connector of the replacement filament to the source interface board.
Figure 69.
Replacing the Filament
Filament
Filament
Retaining
Spring
Screw
Source
Interface
Board