Replacing Components of the Analyzer
Thermo Scientific
TSQ 9000 Hardware Manual
151
.
22. Reattach the Q3 exit lens assembly and Q3 to the analyzer tray. Ensure that quadrupole 3
and the collision cell assembly are correctly positioned before tightening screws.
23. Reattach the two screws on each side of the quadrupole rod endcaps.
Figure 119.
Reattaching Quadrupole 1, the Ion Guide, and the Collision Cell Assembly to the
Analyzer Tray
Figure 120.
Reattaching Quadrupole 1, the Ion Guide, and the Collision Cell Assembly to the
Analyzer Tray
Tip
You will get better signal if you gently push the quadrupole rod endcaps toward one
another while tightening the screws.
Ion Guide
Screws
Q1 Quad
Assembly
Collision Cell
Collision Cell
Assembly
Screws
Q1 Entrance
Lens Endcap
Quadrupole 1