![Thermo Scientific TSQ 9000 Hardware Manual Download Page 165](http://html.mh-extra.com/html/thermo-scientific/tsq-9000/tsq-9000_hardware-manual_1104856165.webp)
3
Advanced Troubleshooting
Replacing Components of the Analyzer
Thermo Scientific
TSQ 9000 Hardware Manual
137
Figure 102.
Finding the Lens Wires between the Quadrupoles
2. Remove the screws on the Q3 exit endcap holding it to the tray.
3. Remove the Q3 quad assembly and Q3 exit lens assembly from the tray.
4. Detach the Q3 exit lens assembly from the Q3 quad assembly.
5. Set the Q3 quad assembly and Q3 exit lens assembly on a clean work surface.
Figure 103.
Removing the Q3 Quad Assembly from the Analyzer Tray
6. Remove the three screws holding the collision cell to the tray.
7. Remove the screw connecting the tray to the collision cell grounding wire. See
.
8. Remove the collision cell assembly and Q1 quad assembly from the analyzer tray.
Lens Wires
CAUTION INSTRUMENT DAMAGE
Quadrupole surface quality is critical for instrument
performance. Take extreme care in handling the quadrupole components.
ATTENTION DOMMAGES À L’INSTRUMENT
La qualité de la surface du quadripôle est
essentielle pour les performances de l’instrument. Soyez extrêmement prudent lorsque vous
manipulez les composants de quadripôle.
Collision Cell
Assembly
Q3 Exit Lens
Assembly
Q3 Quad
Assembly
Screws
Summary of Contents for TSQ 9000
Page 1: ...TSQ 9000 Mass Spectrometer Hardware Manual 1R120618 0003 Revision C December 2018 ...
Page 8: ......
Page 476: ......