Advanced Troubleshooting
Replacing Components of the Manifold
272
TSQ 9000 Hardware Manual
Thermo Scientific
Magnet Yoke, Manifold O-Ring, and Source Retainer Magnets
To replace the magnet yoke, manifold o-ring, or source retainer magnets on the
manifold door
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
. You do not have to remove the ion source cartridge or the column. If you
are using hydrogen as a carrier gas, power off the GC.
2. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Note
If you have an Advanced EI source, follow the steps in
to replace these parts.
Note
For information about replacing the tray alignment pin on the back manifold plate,
Figure 294.
Opening the Manifold Door
Screws