Advanced Troubleshooting
Replacing Components of the Ion Source
Thermo Scientific
TSQ 9000 Hardware Manual
213
6. Insert the ion volume with the handles fitting into the notches of the sleeve. Make sure
the ion volume is firmly seated into the gap on lens 1. You may need to rotate lens 1
slightly to make the ion volume fit correctly.
7. Insert the ion volume repeller insulator onto the repeller.
8. Tighten the repeller nut.
Figure 205.
Inserting the Ion Volume into the Source Sleeve
Note
The ion volume handles are different sizes and will only fit into the sleeve one way.
Ion Volume
Handles
Lens 1 Gap
Handles Fit
Here