9. Use a T20 Torxhead screwdriver to remove the screw connecting the dynode/multiplier
power supply to bottom of the chassis.
10. Pull the dynode power supply straight up and away from the distribution board on the
bottom of the chassis.
11. Replace the dynode/multiplier power supply and install it by aligning its connector with
the connector on the distribution board beneath it. It will drop into place when it is in the
right position.
Figure 388.
Removing the Controller Interface Board Support Bracket
Note
A new dynode/multiplier power supply may have a right angle bracket attached to
the top of the supply. Remove this bracket before placing the supply into the TSQ 9000
instrument.
Dynode/Multiplie
r Power Supply
Screw