Advanced Troubleshooting
Replacing Components of the Analyzer
Thermo Scientific
TSQ 9000 Hardware Manual
149
10. Remove the Q1 entrance lens endcap and ion guide from the tray.
11. Look inside the endcap and use a T10 Torxhead screwdriver to remove the two screws
holding the ion guide to the Q1 entrance lens endcap and Q1 entrance lens.
12. Rotate the ion clamp until the ion guide can be separated from the endcap
13. Pull the ion guide clamp away from the ion guide.
14. Place the ion guide clamp over the new ion guide.
15. Insert the new ion guide into the Q1 entrance lens endcap.
16. Rotate the clamp until the ion guide is locked in place. Make sure the ion guide is aligned
parallel to the flat on the endcap.
17. Tighten the two T10 screws inside the endcap.
Figure 116.
Finding the Quad Entrance Lens
Figure 117.
Replacing the Ion Guide
Note
Make sure the Q1 entrance lens sits flush and centered in the Q1 entrance lens
endcap.
Screws
Ion Guide Wire
Q1 Entrance Lens
Q1 Entrance
Lens Endcap