Advanced Troubleshooting
Replacing Components of the Vacuum Interlock
248
TSQ 9000 Hardware Manual
Thermo Scientific
30. Reattach the two screws connecting the shaft of the vacuum interlock handle to the metal
valve body.
31. Slide the black pillow block over and onto the end of the valve axle.
Note
You need to fully close the sealing ball with the valve axle before you put the handle
back on. Place the handle in the fully down position when you clamp it to the valve axle.
Figure 260.
Reattaching the Metal Valve Body from the Front Manifold Plate
Pillow
Block
Valve Axle
Screws
Vacuum
Interlock
Handle Shaft