3
Advanced Troubleshooting
Replacing Components of the Electron Multiplier
278
TSQ 9000 Hardware Manual
Thermo Scientific
6. Use a T10 Torxhead screwdriver to remove the anode feedthrough, two screws, standoff.
and o-ring. Set them aside on a clean work surface.
7. Use a T20 Torxhead screwdriver to remove the four screws at each corner of the electron
multiplier plate.
Figure 301.
Removing the Anode Feedthrough
Figure 302.
Disconnecting the Electron Multiplier Plate
Anode
Feedthrough
and O-ring
Standoff
Screws
Electron Multiplier
Plate Screws