3
Advanced Troubleshooting
Replacing Components of the Manifold
Thermo Scientific
TSQ 9000 Hardware Manual
271
11. Remove the tray alignment pin and manifold o-ring and put them on the new back
manifold plate. Replace the pin and o-ring, if necessary.
12. Reattach the back manifold plate to the instrument. The back manifold plate must be
reinstalled with the pin at the bottom.
13. Slide the power supply back into the instrument.
14. Reattach the two screws holding the power supply to the instrument.
15. Reconnect the power cables on the left hand sub panel of the TSQ 9000 instrument.
16. Replace the analyzer tray into the instrument by following the instructions in
the Analyzer Tray in the TSQ 9000 Instrument”
17. Reattach the right side panel and top cover panel and tighten the screws.
18. Reattach the GC oven and transfer line.
19. Restart your system by following the instructions in
Figure 293.
Replacing the Back Manifold Plate
Note
Make sure the electrometer board cable is not in the way when you reinstall the
power supply.
Alignment
Pin
Manifold
O-Ring
Back
Manifold
Plate