3
Advanced Troubleshooting
Replacing Components of the Ion Source
196
TSQ 9000 Hardware Manual
Thermo Scientific
10. Reconnect the lens heater block grounding strap to the lens heater block
11. Reconnect the Q1 entrance lens, Q1 exit lens, Q3 entrance lens, and Q3 exit lens to the
source interface board using a flat-tipped screwdriver.
12. Once the components are correctly reattached, follow the instructions in
Analyzer Tray in the TSQ 9000 Instrument”
to replace the tray in the TSQ
instrument.
13. Restart your system by following the instructions in
Figure 182.
Reconnecting the Lens Heater Block Grounding Strap
Lens
Heater
Block
Lens Heater
Grounding Strap
Screw