Performing Routine Maintenance
Removing the Ion Source Cartridge for Cleaning
46
TSQ 9000 Hardware Manual
Thermo Scientific
a. With your hand around the stainless steel barrel, twist the tool until the alignment
groove at the end of the barrel aligns with the left screw around the slot vacated by
the vacuum interlock knob.
b. Twist and push the stainless steel barrel to the right until it clicks into place under the
screw at the top of the slot.
Figure 33.
Inserting the Ion Volume Tool
Figure 34.
Twisting the Tool Into Place