3
Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
330
TSQ 9000 Hardware Manual
Thermo Scientific
21. Reinsert the lens driver board by aligning the connector on the bottom of the board with
the connectors on the distribution board beneath it. It will drop into the slot when it is in
the right position.
22. Push the board down into the connector.
23. Look through the left side of the instrument and tighten the screw that holds the lens
driver board in place on the chassis.
24. Reattach the top cover panel and tighten the screw.
25. Restart your system by following the instructions in
Figure 354.
Reinstalling the Lens Driver Board
Connector for
Lens Driver
Board