3
Advanced Troubleshooting
Replacing the Calibration Gas Components
374
TSQ 9000 Hardware Manual
Thermo Scientific
26. Reinstall the lens driver board by aligning the connector on the bottom of the board with
the connectors on the distribution board beneath it. It will drop into the slot when it is in
the right position.
27. Push the board down into the connectors.
28. Look through the left side of the instrument and use a T20 Torxhead screwdriver to
reattach the screw holding the lens driver board in place on the chassis.
Note
This screw is required for good ground contact.
Figure 398.
Reinstalling the Lens Driver Board
Connector for
Lens Driver
Board