3
Advanced Troubleshooting
Replacing Components of the Analyzer
Thermo Scientific
TSQ 9000 Hardware Manual
147
Ion Guide
To replace the ion guide:
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the TSQ 9000 Instrument”
3. Turn the analyzer tray to access the bottom and disconnect the three lens wires attached
to the lenses between the quadrupoles.
Figure 112.
Finding the Lens Wires between the Quadrupoles
4. Remove the screws holding the Q3 exit lens endcap to the tray.
5. Remove Q3 and the Q3 exit lens endcap.
6. Remove the screws holding the collision cell to the tray.
7. Remove the collision cell and Q1.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Lens Wires
CAUTION INSTRUMENT DAMAGE
Quadrupole surface quality is critical for instrument
performance. Take extreme care in handling the quadrupole components.
ATTENTION DOMMAGES À L’INSTRUMENT
La qualité de la surface du quadripôle est
essentielle pour les performances de l’instrument. Soyez extrêmement prudent lorsque vous
manipulez les composants de quadripôle.