Advanced Troubleshooting
Replacing Components of the Ion Source
202
TSQ 9000 Hardware Manual
Thermo Scientific
5. Evacuate the VPI.
a. Confirm that the source removal tool is properly engaged in the VPI.
b. Press the blue
Evacuate
button on the front of the instrument.
c.
The Evacuate light will begin to flash green and should continue to flash green for
approximately 20 seconds.
d. If the pressure has returned to an acceptable value after the 20-30 second wait, the
evacuate light will turn off and the
Ready to Open
light will be solid green. At that
point, the air has been evacuated from the ball valve and it is safe to open the vacuum
interlock valve.
6. Twist, then loosen the top part of the vacuum interlock handle to allow the entire handle
to swing up and down. This handle controls the interior ball valve, which seals the
vacuum chamber.
Note
If the evacuate light flashes for only a short time then returns to a solid light
without the Ready to Open light turning green, there is a leak in the VPI seal. The
source removal tool connection should be checked or field service contacted.
Figure 189.
Evacuating the Instrument
Evacuate
Button
Vacuum
Interlock
Handle