3
Advanced Troubleshooting
Replacing Components of the Ion Source
Thermo Scientific
TSQ 9000 Hardware Manual
193
4. Use a T10 Torxhead screwdriver to disconnect the lens heater block grounding strap from
the lens heater block, but leave it attached to the tray.
5. Use a T20 Torxhead screwdriver to remove the two screws on the back of source block.
Figure 176.
Disconnecting the Lens Heater Block Grounding Strap
Figure 177.
Disconnecting the Source Block to the Analyzer Tray
Note
Some of the pieces may detach as you loosen the screws. Do not let them fall.
Lens
Heater
Block
Lens Heater
Grounding Strap
Screw
Screws