![Thermo Scientific TSQ 9000 Hardware Manual Download Page 208](http://html.mh-extra.com/html/thermo-scientific/tsq-9000/tsq-9000_hardware-manual_1104856208.webp)
3
Advanced Troubleshooting
Replacing Components of the Ion Source
180
TSQ 9000 Hardware Manual
Thermo Scientific
5. Extract the thumbscrew-repeller spacers and transfer to new thumbscrews.
6. Attach new source-repeller thumbscrews to the ion source.
7. Reattach the manifold door and the four screws.
8. Restart your system by following the instructions in
Figure 160.
Replacing the Source-Repeller Thumbscrews
Source-Repeller
Thumbscrews
Summary of Contents for TSQ 9000
Page 1: ...TSQ 9000 Mass Spectrometer Hardware Manual 1R120618 0003 Revision C December 2018 ...
Page 8: ......
Page 476: ......