Advanced Troubleshooting
Replacing Components of the Ion Source
Thermo Scientific
TSQ 9000 Hardware Manual
229
5. Twist the vent valve knob one and a half times in a counter-clockwise direction to open
the vent.
6. Wait five minutes for venting to complete.
7. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door. See
.
Figure 225.
Venting the TSQ 9000 Mass Spectrometer
Vent Valve Knob
CAUTION - INSTRUMENT DAMAGE
Do not proceed until the instrument is vented, or
pieces of the column or ferrule might blow into the instrument. To ensure that the
instrument is vented, check how much the glass cover compresses the top cover o-ring in
the manifold. Once the o-ring surface touching the glass is about 1 mm, it is safe to open
the instrument and remove the column.
Note
Remove the screws completely so that they do not scratch the manifold when you
close the door.