3
Advanced Troubleshooting
Replacing Components of the Manifold
Thermo Scientific
TSQ 9000 Hardware Manual
269
Back Manifold Plate
To replace the back manifold plate:
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Pull the column fully out of the transfer line and move the instrument away from the GC.
3. Slide the top cover panel toward the back of the instrument and lift it off.
4. Use a T20 Torxhead screwdriver to remove the screw at each corner of the right side
panel.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge.
Figure 291.
Removing the Right Side Panel Screws
Right Side
Panel Screws