Replacing Components of the Ion Source
182
TSQ 9000 Hardware Manual
Thermo Scientific
4. Twist off each source-repeller thumbscrew one at a time to prevent the repeller plate from
falling out.
5. Remove the thumbscrew-repeller spacers from the thumbscrews and replace them.
6. Reattach the source-repeller thumbscrews to the spacers.
7. Reattach the manifold door and the four screws.
8. Restart your system by following the instructions in
Figure 162.
Replacing the Source-Repeller Thumbscrews
Figure 163.
Replacing the Thumbscrew-Repeller Spacers
Note
When reattaching the spacers, make sure they fit into the grooves on the
thumbscrews. Also, do not over-tighten the thumbscrews or it may crack the spacers.
Source-Repeller
Thumbscrews
Thumbscrew-Repeller
Spacers