3
Advanced Troubleshooting
Replacing Components of the Ion Source
Thermo Scientific
TSQ 9000 Hardware Manual
223
8. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door. See
.
9. Open the manifold door. You can now see the ion source block.
Figure 218.
Locating the Ion Source Block
10. Insert the small source removal tool into the ion source block as shown in
.
Note
Remove the screws completely so that they do not scratch the manifold when you
close the door.
Figure 217.
Removing the Manifold Door Screws
Ion Source Block