Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
Thermo Scientific
TSQ 9000 Hardware Manual
351
Source Interface Board and Heat Shield
To replace the source interface board and heat shield
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. If you are only replacing the heat shield, you do not need to remove the analyzer tray.
Disconnect the filament board and wires from the source interface board and the left
screw to remove the heat shield.
3. Replace the heat shield.
4. Reconnect the filament board and wires to the source interface board.
5. If you are replacing the source interface board, remove the analyzer tray from the
instrument by following the instructions in
“Removing the Analyzer Tray from the TSQ
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components.
Figure 377.
Replacing the Heat Shield
Note
Keep all manifold parts you remove covered on a clean work surface so they will not
accumulate dust.
Source Interface
Board
Heat Shield
Left
Screw
Filament Board
and Wires