Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
Thermo Scientific
TSQ 9000 Hardware Manual
325
7. Use your fingers to pry the bracket of the board away from the chassis and toward the left
side of the instrument.
8. Remove the top manifold cover (glass) out of the top of the instrument and set it aside.
9. Open the front door of the TSQ 9000 instrument.
10. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Figure 346.
Removing the Lens Driver Board
Note
Remove the screws completely so that they do not scratch the manifold when you
close the door.
Lens Driver Board Screw