3
Advanced Troubleshooting
Replacing Components of the Analyzer
Thermo Scientific
TSQ 9000 Hardware Manual
159
Ion Guide Wire Set
To replace the ion guide wires
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the TSQ 9000 Instrument”
3. Use a T10 Torxhead screwdriver to remove the screw connecting the ion guide wire to the
top of the ion guide.
4. Use a wrench to remove the nut connecting the other ion guide wire to the side of the ion
guide.
5. Attach new ion guide wires and reattach the nuts and screw.
6. Follow the instructions in
“Replacing the Analyzer Tray in the TSQ 9000 Instrument”
to replace the tray into the instrument.
7. Restart your system by following the instructions in
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components.
Figure 131.
Replacing the Ion Guide Wires
Screw
Nut
Ion Guide Wires