3
Advanced Troubleshooting
Replacing Components of the Vacuum Interlock
Thermo Scientific
TSQ 9000 Hardware Manual
245
20. Remove the axle bearing from the metal valve body. Replace the axle bearing, if necessary.
21. Pull the valve axle out of the side of the metal valve body. Replace the valve axle, if
necessary.
22. Remove the axle seal from the metal valve body. Replace the axle seal on the valve axle if
necessary.
Figure 254.
Replacing the Axle Bearing
Figure 255.
Replacing the Axle Seal
Axle Bearing
Clip
Safety
Interlock
Axle Seal