3
Advanced Troubleshooting
Replacing Components of the Ion Source
Thermo Scientific
TSQ 9000 Hardware Manual
173
12. Use a T20 Torxhead screwdriver to remove the two screws on the back of the source
heater block.
13. Slide the source interface board and the source block out from under the tray without
damaging or disconnecting the wires.
Figure 148.
Disconnecting the Ion Source to the Analyzer Tray
Note
Some of the pieces may become slack as you loosen the screws. Do not let them fall.
Screws
CAUTION - INSTRUMENT DAMAGE
The lens/source heater may be fused to the source
block and difficult to remove. To avoid damage, separate the block from the tray before
removing the lens/source heater.
ATTENTION DOMMAGES À L’INSTRUMENT
Le chauffage de lentille/source peut être
fusionné au bloc source et difficile à enlever. Pour éviter les dommages, séparez le bloc du
plateau avant d’enlever le chauffage de lentille/source.
Figure 149.
Removing the Ion Source from the Analyzer Tray
Wires
Ion Source
Block