3
Advanced Troubleshooting
Replacing Components of the Analyzer
Thermo Scientific
TSQ 9000 Hardware Manual
131
Figure 92.
Disconnecting the Ion Guide Connector Wires and the Quad 1 Connector Wires
6. Disconnect the four wires that go to the collision cell and quadrupole 3. These wires are
attached to the rear 4-pin feedthrough in the vacuum manifold.
Figure 93.
Disconnecting the Quad 3 Connector Wires and the Collision Cell Connector Wires
7. Remove the collision gas tube from the collision cell.
8. Use a T20 Torxhead screwdriver to remove the four screws around the manifold door.
Note
Pull straight back from the collision cell to prevent damage to the collision gas tube.
Note
Remove the screws completely so that they do not scratch the manifold when you
close the door.
Quad 1
Wires
Front 4-pin
Connector
Ion Guide Wires
Quad 3
Wires
Rear 4-pin
Connector
Collision Cell
Connector Wires
Collision Gas
Tube