Advanced Troubleshooting
Replacing the Boards in the TSQ 9000 Instrument
Thermo Scientific
TSQ 9000 Hardware Manual
353
Electrometer Board Cable
To replace the electrometer board (15-pin male/female RS-232) cable
1. Shut down your system by following the steps in
“Shutting Down the TSQ 9000 System”
. You do not have to remove the ion source cartridge or the column or vent
the instrument. If you are using hydrogen as a carrier gas, power off the GC.
2. Slide the top cover panel toward the back of the instrument and lift it off.
3. Disconnect the electrometer board cable
4. Remove all power connections from the TSQ 9000 instrument.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Figure 379.
Disconnecting the Electrometer Board Cable
Electrometer
Board Cable