1
Performing Routine Maintenance
Configuring the TSQ 9000 System
Thermo Scientific
TSQ 9000 Hardware Manual
5
2. Connect all the cables to the computer, keyboard, mouse, monitor, and printer (if
applicable).
3. Connect the handshaking cable to the GC and to the TSQ 9000 instrument’s GC Start
connector.
4. Connect the computer and monitor electrical cables to the wall outlet.
5. Connect the GC communication cable to the COM1 or LAN port on the back of the
computer.
6. Connect the Ethernet cable to the Data System connector on the TSQ 9000 instrument
and to the bottom Ethernet port on the back of the computer.
7. Power-on the computer, monitor, and printer.
8. Set the time, date, and passwords.
9. Install print drivers and test the printer.
10. Connect the foreline vacuum hose to the foreline pump input using the o-ring screen seal
and clamp.
11. Connect the exit of the foreline pump to a suitable exhaust or oil mist filter.
12. Connect the foreline pump’s power cable to the pump and to the System Mech Pump
connector on the TSQ 9000 instrument.
13. Turn the foreline pump power switch to the
On
position. The foreline pump will not
power on until the TSQ 9000 instrument is powered on.
Note
Leave at least 46 cm (18 in.) of free space on the left side of the instrument to
provide easy access to the power switch and other appliance inlets.
Note
If you are using a probe controller, connect the probe controller cable to the
Accessory Start connector on the left hand sub panel of the TSQ 9000 instrument.
CAUTION
-
INSTRUMENT DAMAGE:
Be sure to connect the foreline vacuum hose to the
foreline pump inlet on the roughing pump. If you attach the hose to the wrong inlet on
the roughing pump, pump oil will flood the manifold.
CAUTION
INSTRUMENT DAMAGE:
Make sure the voltage on the pump matches the
AC voltage that is powering the TSQ 9000 system.
ATTENTION DOMMAGES À L’INSTRUMENT
: assurez-vous que la tension au niveau de la
pompe correspond à la tension CA qui alimente l’instrument TSQ 9000.