APPLIED OPERATIONS
/Image Acquisition
IV.
OPERATION INSTRUCTIONS
IV .
2 - 5 6
Page
2-2-3 Differences in Image Acquisition Method Between Fluorescent and
Transmitted Images
2-2-3-1 Monochrome Image
The wavelength obtained by monochrome dyeing can be acquired and observed as an
image of a channel (either Ch1 or Ch2).
The subsequent description deals with the differences in operation between the
monochrome dyeing and dual-fluorochrome dyeing.
When the specimen is dyed so that the wavelength of the fluorescence emission from it
is
below 570 nm
, the image is acquired in
channel 1 (Ch1)
.
When the specimen is dyed so that the wavelength of fluorescence emission from it is
above 570 nm
, the image is acquired in
channel 2 (Ch2)
.
z
Set the [DETECTION MODE] lever of the scan unit to “FL -” or “- FL”.
FL -:
When acquiring the image in Ch1.
- FL:
When acquiring the image in Ch2.
z
Engage the filter by setting the [BARRIER FILTERS] lever of the scan unit. (Setting
the lever to “IN” engages the filter.)
CH1:
When observing a specimen dyed to generate fluorescence wavelength
below 570 nm.
CH2:
When observing a specimen dyed to generate fluorescence wavelength
above 570 nm.
Also see section 2-2-1-2, “Configuring the Scan Unit“ and follow the instructions given
by the [Microscope Configuration] window.
Summary of Contents for FLUOVIEW FV300
Page 2: ......
Page 8: ......
Page 10: ......
Page 12: ......
Page 26: ......
Page 28: ......
Page 38: ......
Page 114: ......
Page 486: ......
Page 506: ......
Page 512: ......
Page 518: ......
Page 520: ......
Page 548: ......
Page 552: ......
Page 554: ......
Page 556: ......
Page 562: ......
Page 563: ......