APPLIED OPERATIONS
/Image Acquisition
IV.
OPERATION INSTRUCTIONS
IV .
2 - 8 5
Page
2-2-10 Display the change of image intensity (Point Scan)
The change of fluorescence intensity according to time-lapse can be displayed
graphically by irradiating the laser to certain point on the image.
If you have FLUOVIEW TIME COURSE software, image acquisition can be started with
buttons, keys or input of external trigger signals.
1. Acquire an image in the XY observation mode. For the operation procedure, see
section 2-2-1, “Image Acquisition with XY observation”.
2. Select <Point> under the [Point] option button in the [Mode] group box in the
[Acquire] panel.
Shows the X, Y
coordinates on the
image.