TROUBLESHOOTING GUIDE
VI .
TROUBLESHOOTING
VI.
1 - 2
Page
Phenomenon
Cause
Treatment
Manual Ref. Pages
1. Fluorescence image
cannot be observed
The fluorescent dye does not
match the barrier filter.
Select the barrier filter
matching the fluorescent
dye.
See 2-2-1-2, “Configuring
the Scan Unit” in
OPERATION
INSTRUCTIONS and follow
instructions of the
[Microscope Configuration]
window.
2-2-2-1,
OPERATION
INSTRUCTIONS
1-2-8 & 1-3-2,
OPERATION
INSTRUCTIONS
Focusing is not correct.
Adjust focusing.
1-2-3, OPERATION
INSTRUCTIONS
The photomultiplier tube
voltage of the detector is low.
Increase the photomultiplier
tube voltage using the [PMT]
scale in the [Acquire] panel.
2-2-1-3-9,
OPERATION
INSTRUCTIONS
The OFFSET value is too
high.
Decrease OFFSET to an
optimum value.
2-2-1-3-9,
OPERATION
INSTRUCTIONS
The photomultiplier tube of
the channel set with the
detection mode setting slider
is not set.
Check the [Ch1], [Ch2] or
[Ch3] check box in the
[Acquire] panel.
1-1-1 & 2-2-1-3-3,
OPERATION
INSTRUCTIONS
2. Transmitted image cannot
be observed.
The transmittance detection
channel is not selected.
In the [Acquire] panel, check
the check box of the [Ch]
with which “Transmitted” is
displayed.
2-2-3-3,
OPERATION
INSTRUCTIONS
The transmitted light filter is
engaged in the light path of
the microscope.
Disengage the filter from the
light path.
1-1-2 & 1-2-4,
OPERATION
INSTRUCTIONS
3. Image is disturbed.
The installation location is
subject to large vibrations.
Consult Olympus.
-
Darken the room before
acquiring image.
-
External light from a
fluorescent lamp, etc. is
detected.
Ensure that the top cover of
the scan unit is closed tightly.
1-1-1, OPERATION
INSTRUCTIONS
4. Image is striped.
The laser reflection light is
penetrating.
The barrier filter selection is
erroneous.
1-2-8, OPERATION
INSTRUCTIONS
5. Image looks poor.
The analyzer is engaged in
the light path.
Disengage the analyzer from
the light path.
1-2-3 & 1-2-4,
OPERATION
INSTRUCTIONS
The scan speed is high.
Decrease the scan speed to
an optimum speed or
integrate the image.
2-2-1-3-10 & 2-2-1-
5, OPERATION
INSTRUCTIONS
Summary of Contents for FLUOVIEW FV300
Page 2: ......
Page 8: ......
Page 10: ......
Page 12: ......
Page 26: ......
Page 28: ......
Page 38: ......
Page 114: ......
Page 486: ......
Page 506: ......
Page 512: ......
Page 518: ......
Page 520: ......
Page 548: ......
Page 552: ......
Page 554: ......
Page 556: ......
Page 562: ......
Page 563: ......