APPLIED OPERATIONS
/Image Acquisition
IV.
OPERATION INSTRUCTIONS
IV .
2 - 1 0 7
Page
2-2-13 Notes for image acqui sition
2-2-13-1 Memory of setting information for scanning region
Frame with handle described in 2-2-5 Clip Scan, Frame with hand described in 2-2-6
Zoom-in Scan, Straight line with handle described in 2-2-9, Cross line cursor described
in 2-2-10 Point Scan are all called Scanning Region Setting - ROI (Region Of Interest).
This Scanning Region Setting – ROI holds the setting information described below when
scanning is executed; and the information appears in the same state when each scan
mode is selected next time.
·Angle
·Profile
·Position
When default setting information is desired,
click (mouse right) over circle region in [Pan Zoom] group box.
As [Reset current scan ROI] menu appears, select it.
The ROI selected at this time is returned to default setting information.
When size of image acquisition is changed, all ROI setting information
would be initialized.
2-2-13-2 Limits for image acquisi tion
Depending upon the system used, some scanning modes cannot be used. It
depends upon kind of I/O board and kind of combiner, etc. Table for combination is
shown below.
ROI
䌈䌡䌮䌤䌬䌥
ᵈᗧ