Getting Started FLUOVIEW/
Outline of LSM Observation Procedures
IV.
OPERATION INSTRUCTIONS
IV.
1 - 2 5
Page
Change right side rotary switch only. Do not touch left side rotary
switch. If you change left rotary switch, you may not be able to get
optimum performance of this system.
When you use LD405, align arrow mark of right side rotary switch with
5.
When you use LD440, align arrow mark of right side rotary switch with
4.
To stabilize the laser beam output, we recommend leaving the system
for warm-up after turning the laser power ON. The warm-up period
should exceed 10 minutes when the laser system in use uses an Ar
laser or exceed 30 minutes when it uses a Kr laser, Green HeNe laser
or HeCd laser.
x
Turning the reflected light power supply unit
(1) Set the power switch to ON.
Once the reflected light power supply unit is turned OFF, do not turn it
ON again for at least 10 minutes. Otherwise the service life of the
mercury burner will be shortened.
x
Turning the monitor ON
(Refer to the instruction manual of your monitor.)
As this unit should be connected to the power outlet unit, leave its
power switch in the ON position.
x
Turning the microscope ON
(Refer to the instruction manual of the microscope combined with the
system.)
As this unit should be connected to the power outlet unit, leave its
power switch in the ON position.
x
Turning the transmitted light unit (when the transmitted light detector is
used)
NOTE
(1)
Rotary switch
(In case of LD405)
(In case of LD440)
Summary of Contents for FLUOVIEW FV300
Page 2: ......
Page 8: ......
Page 10: ......
Page 12: ......
Page 26: ......
Page 28: ......
Page 38: ......
Page 114: ......
Page 486: ......
Page 506: ......
Page 512: ......
Page 518: ......
Page 520: ......
Page 548: ......
Page 552: ......
Page 554: ......
Page 556: ......
Page 562: ......
Page 563: ......