Getting Started FLUOVIEW/
Outline of LSM Observation Procedures
IV.
OPERATION INSTRUCTIONS
IV.
1 - 3 3
Page
1-2-4-2 Combination with Erected Microscope(AX)
1. Pull out the light path selector (1) on the trinocular eyepiece fully to the stop
position.
2. Pull out the light path selector (7) on the TV adapter fully to the stop position.
3. Select the cube in the cube cassette (2) for the vertical illuminator by pressing the
required cube conversion button (8) on the U-HSTR hand switch.
4. When the U-AN analyzer (3) is in use, disengage it from the light path by
setting the switch to the pulled-out position.
5. When only fluorescence observation is required, disengage the U-DICR
transmitted DIC slider (4) by setting the switch to the pulled-out position. When
transmitted light differential interference observation or simultaneous fluorescence
+ transmitted light differential interference observation is required, engage the U-
DICT and the optimum transmitted light DIC slider for the objective in the light path
by operating the universal condenser (6).
6. For transmitted light observation, disengage any filter (5) from the light path.
(2) Cube cassette for
vertical illuminator
AX-URBC
(3) Analyzer
AX-AN (optional)
(4) Transmitted DIC slider
U-DICR(optional)
(5) Filters
Filter lever (ND)
Filter levers (IF550, LBD)
(1) Light path selector knob
BI
LSM
(6) Universal condenser
Hand switch for transmitted reflectance
U-HSTR
(7) Light path selector knob
TV,Photo
LSM
(8)
Summary of Contents for FLUOVIEW FV300
Page 2: ......
Page 8: ......
Page 10: ......
Page 12: ......
Page 26: ......
Page 28: ......
Page 38: ......
Page 114: ......
Page 486: ......
Page 506: ......
Page 512: ......
Page 518: ......
Page 520: ......
Page 548: ......
Page 552: ......
Page 554: ......
Page 556: ......
Page 562: ......
Page 563: ......