The accumulation count can be set up a maximum of 63 times.
When 0 is set as the number of times of accumulation, an ordinary image
acquisition is to be performed.
4.
Click the <Once> button in the [Acquire] panel.
The acquired image will be displayed in the [Live] panel.
And the accumulated image can be displayed in the [Display] panel without being
displayed in full.
The difference between the Frame and Line modes of Kalman Accumulation
In frame mode, the image is accumulated every one frame during acquisition.
In line mode, accumulation is done every one line during acquisition.
The frame mode is intended to utilize for fixed specimen observation, and the line mode is for
living specimen observation. Line mode makes it possible to shorten sampling interval so
that it is suitable for live cell observation.
2-2-1-6 Saving the Acquired Ima ge in File
1. Display the [File I/O] panel.
2. Click the page tab of the [Live] panel showing the image to be saved, so that the
image is displayed at the front.
3. Click the <Experiment> button in the [Save] group box in the [File I/O] panel.
For details, see section 2-3-1, “Saving Image”.